Ultra-High-Resolution UV–Vis–NIR Fiber-Optic Spectrometer
Model GR400P
Uses a room-temperature 4,096-pixel sensor and a 16-bit ADC to measure ultraviolet, visible, and near-infrared spectra.

Product Overview
The GR400P has four typical grating configurations. The 300@dual-blaze configuration covers 200–1100 nm. The 1200@1000 configuration covers 810–980 nm and achieves an optical bandwidth of less than 0.3 nm with a 10 μm slit.
Technical Specifications
- Detector
- 4,096-pixel room-temperature CMOS linear array
- ADC Resolution
- 16-bit
- Wavelength Accuracy
- ≤0.5nm
- Stray Light
- <0.1%
- Signal-to-Noise Ratio
- 350:1
- Dynamic Range
- 3500:1
- Integration Time
- 108us - 65s
- Power Supply
- 5 V DC / 250 mA (USB bus power) or 12 V DC / 1 A (DB-15)
- Communication Interface
- USB-to-serial (STM32 virtual COM port), RS232 (DB-15)
- Operating Temperature
- -5℃–55℃
- Dimensions
- 150×115×45mm/900g
Typical Configurations (Measured Values Vary by Target Wavelength. Custom Configurations Available.)
| GR400P | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| No. | Grating (grooves/mm@nm) | Start Wavelength (nm) | End Wavelength (nm) | Spectral Width (nm) | Sampling Interval (nm/pixel) | Slit Width (μm) | ||||
| 10 | 25 | 50 | 100 | 200 | ||||||
| Optical Bandwidth (<nm) | ||||||||||
| 1 | 300@dual blaze | 200 | 1100 | 900 | 0.22 | 2 | 2.3 | 2.5 | 4.8 | 7.5 |
| 2 | 400@500 | 190 | 850 | 660 | 0.16 | - | - | - | 3.4 | - |
| 3 | 900@800 | 850 | 1079 | 229 | 0.06 | 0.4 | - | - | - | - |
| 4 | 1200@1000 | 810 | 980 | 170 | 0.04 | 0.3 | - | - | - | - |
Applications
- UV, visible, and near-infrared spectral measurement