High-Resolution UV–Vis–NIR Fiber-Optic Spectrometer
Model GR400
Measures UV–Vis–NIR spectra for fluorescence, transmittance and reflectance. Grating and slit options can be selected for the target wavelength range.

Product Overview
The GR400 series uses a crossed Czerny–Turner optical design. The grating, cylindrical mirror, and filter are selected for the target wavelength range, delivering high spectral resolution and a high signal-to-noise ratio. A standard industrial USB 2.0 Type-B port provides power and data transfer for most spectroscopy applications. The spectrometer communicates over an STM32 virtual COM port and works without an additional driver on most operating systems. A D-Sub connector supports external power, I/O, and triggering with a fully wired VGA cable. Double-Briiiant Technology also provides QuantumSpectX, a cross-platform spectroscopy application written in C++ for quickly acquiring spectra and viewing instrument information, and a C++ communications SDK for custom development.
Product Images
Technical Specifications
- Detector
- 2,048-pixel room-temperature CMOS linear array
- ADC Resolution
- 16-bit
- Wavelength Accuracy
- ≤0.5nm
- Stray Light
- <0.1%
- Signal-to-Noise Ratio
- 350:1
- Dynamic Range
- 3500:1
- Integration Time
- 108us - 65s
- Power Supply
- 5 V DC / 250 mA (USB bus power) or 12 V DC / 1 A (DB-15)
- Communication Interface
- USB-to-serial (STM32 virtual COM port), RS232 (DB-15)
- Operating Temperature
- -5℃–55℃
- Dimensions
- 150×115×45mm/900g
Typical Configurations (Measured Values Vary by Target Wavelength. Custom Configurations Available.)
| GR400 | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| No. | Grating (grooves/mm@nm) | Start Wavelength (nm) | End Wavelength (nm) | Spectral Width (nm) | Sampling Interval (nm/pixel) | Slit Width (μm) | ||||
| 10 | 25 | 50 | 100 | 200 | ||||||
| Optical Bandwidth (<nm) | ||||||||||
| 1 | 300@dual blaze | 200 | 1100 | 900 | 0.44 | 2 | 2.3 | 2.5 | 4.8 | 7.5 |
| 2 | 400@500 | 300 | 1000 | 700 | 0.34 | - | 1.3 | 2.3 | 2.7 | - |
| 3 | 500@560 | 300 | 845 | 545 | 0.27 | 0.7 | - | - | - | - |
| 4 | 600@500 | 229 | 900 | 671 | 0.33 | - | 2.2 | - | - | - |
| 5 | 1200@640 | 700 | 855 | 155 | 0.08 | 0.6 | - | - | - | - |
| 6 | 1800@640 | 650 | 840 | 190 | 0.09 | - | - | 0.8 | - | - |
| 7 | 2400@500 | 425 | 505 | 80 | 0.04 | 0.25 | - | - | - | - |
Sample Factory Test Report
Measured on a GR400 sample unit. The report includes the optical configuration, measured optical bandwidth, and a reference spectrum.
Download Report
Applications
- UV, visible, and near-infrared spectral measurement
- Fluorescence measurement
- Transmittance and reflectance measurement
- Absorbance measurements under high-intensity illumination
- Laser and LED characterization
- CIE colorimetry