Miniature High-Resolution Fiber-Optic Spectrometer
Model GR200
Uses a room-temperature 2,048-pixel sensor and a 16-bit ADC to measure ultraviolet, visible, and near-infrared spectra.

Product Overview
A typical GR200 configuration uses a 600@500 grating and covers 250–900 nm. With a 25 μm slit, the optical bandwidth is less than 1.5 nm.
Product Images
Technical Specifications
- Detector
- 2,048-pixel room-temperature CMOS linear array
- ADC Resolution
- 16-bit
- Wavelength Accuracy
- ≤0.5nm
- Stray Light
- <0.1%
- Signal-to-Noise Ratio
- 350:1
- Dynamic Range
- 3500:1
- Integration Time
- 108us - 65s
- Power Supply
- 5 V DC / 250 mA (USB bus power) or 6–9 V DC / 1 A (2 × 5 pin, 1.27 mm pitch, square-hole female header)
- Communication Interface
- USB-to-serial (STM32 virtual COM port), RS232 (2 × 5 pin, 1.27 mm pitch, square-hole female header)
- Operating Temperature
- -5℃–55℃
- Dimensions / Weight
- 90*65*30mm/400g
Typical Configurations (Measured Values Vary by Target Wavelength. Custom Configurations Available.)
| GR200 | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| No. | Grating (grooves/mm@nm) | Start Wavelength (nm) | End Wavelength (nm) | Spectral Width (nm) | Sampling Interval (nm/pixel) | Slit Width (μm) | ||||
| 10 | 25 | 50 | 100 | 200 | ||||||
| Optical Bandwidth (<nm) | ||||||||||
| 1 | 600@500 | 250 | 900 | 650 | 2.54 | - | 1.5 | - | - | - |
Sample Factory Test Report
Measured on a GR200 sample unit. The report includes the optical configuration, measured optical bandwidth, and a reference spectrum.
Download Report
Applications
- UV, visible, and near-infrared spectral measurement